Dissolved oxygen analyzer ADS550
Accurate and reliable measurement of dissolved oxygen in high purity water.
The Navigator 500 dissolved oxygen analyzer provides continuous monitoring and control of dissolved oxygen levels in steam-raising and desalination plants and semiconductor water treatment.
- Measurement range
0 to 20,000 ppb - Measurement method
Continuous dissolved oxygen measurement using electrochemical cell - Accuracy / maximum measured error
±5 % of reading or ±1 ppb, whichever is the greater - Repeatability
±3 % of reading or ±1 ppb, whichever is the greater - Response time
1 minute for a 90 % step change - Resolution
0.1 ppb - Calibration
Automatic calibration